    Automated Protein Quantification with the Biomek-FX Liquid Handler System

    Yan Chen, Nurgul Kaplan Lease, Jennifer Gin, Tad Ogorzalek, Christopher J Petzold

    Published: 2022-01-13 DOI: 10.17504/protocols.io.b3grqjv6
    AI 解读
    浏览 1525

    Abstract

    This protocol details steps to perform the protein quantification (Lowry-based) assay by using a Biomek FX liquid handler system. It is optimized to assay a full 96-well plate of protein samples in duplicate with a separate (control) plate for BSA standards. You will need a plate reader to measure the samples and standards.

    This protocol works best as part of a full proteomic sample preparation workflow with:

    Automated Chloroform-Methanol Protein Extraction on the Biomek-FX Liquid Handler System

    and

    Automated Protein Normalization and Tryptic Digestion on a Biomek-FX Liquid Handler System

    Before start

    Prepare BSA Standards Plate (1st 4 rows from A to D):

    1. Add 40 uL of H2O into wells A1 to D1.

    2. Add 40 uL of BSA Standards 1 (125 ug/mL) to 7 (2000 ug/mL) into columns 2 to 8. For this protocol you will need:

    3. Beckman-Coulter Biomek FX liquid handler system with a 96-pod head.

    4. Upload the attached method file and modify it to fit your deck and system configuration. Modular Protein Quantitation method.bmf

    Steps

    Deck Setup

    1.

    Open Biomek Software that controls Biomek-FX liquid handler system. Under "File" drop-down menu, click "Open" to select the automation method "Modular Protein Quantitation method."

    Note
    Because different deck orientations and system components are possible, you will need to modify the method file (attached in the 'Before start' section) for your specific Biomek liquid handler system.

    2.

    Click on "Instrument Setup" under the "Setup" group node to get visual instruction on how to set up the deck.

    3.

    Set up the deck (refer to the deck setup picture below):

    ABC
    Deck LabelLabwareReagent
    proteinPCR96 plate (BIO-RAD, Cat.#HSP9601)unknown amount of protein to quantify
    titratePCR96 plate (BIO-RAD, Cat.#HSP9601)
    BSA standardsPCR96 plateBSA Standards (Thermo Fisher, Cat.#23208)
    tips1,220 µl pipet tips (Molecular Bioproducts BioRobotix, Cat.#918-262 )
    tips 3,5200 uL pipet tips (Molecular Bioproducts BioRobotix, Cat.#919-262 )
    control read, prot read 1, prot read 2Black Microplate (Fisher Scientific, Cat.#07-200-567)
    Buffer AReservoir Microplate (Agilent, Cat.#201254-100)DC Protein Assay Reagent A (Bio-rad Laboratories, Cat.#500-0113)
    waterReservoir Microplate (Agilent, Cat.#201254-100)LC-MS grade Water (VWR Scientific, Cat.#BJLC365-2.5)
    Buffer B96 Deep Well Reagent Reservoir (VWR, Cat.#101100-962)DC Protein Assay Reagent B (Bio-rad Laboratories, Cat.#500-0114)

    Materials for Deck setup

    Deck setup
    Deck setup
    Labware for Deck setup
    Labware for Deck setup
    4.

    MANUAL STEP: Use a multichannel pipette to mix protein samples right before starting.

    5.

    Click the "Run" button (green arrow) to start.

    DC protein assay

    6.

    Transfer 25 µl of Buffer A to Protein Read plate 1, Protein Read plate 2, and Control Read plate.

    7.

    Transfer 12 µl of H2O into Titrate plate. Then transfer 3 µl (see Note for more details) from Protein plate to Titrate plate and mix with 5 cycles of pipetting mixing on deck.

    Note
    Note: The dilution factor could be altered as needed by changing the volumes of water and proteins transferred to the titration plate. Be sure to multiply the protein quant results by the dilution factor before you do your normalization calculation.

    8.

    Transfer 5 µl of protein from Titrate plate to Protein Read plate 1 and Protein Read plate 2.

    9.

    Transfer 5 µl from BSA Standards plate to Control Read plate.

    Note
    Prepare BSA Standards plate (1st 4 rows from A to D):Add 40 µl of H2O into wells A1 to D1.Add 40 µl of BSA Standards 1 (125 µg/ml) to 7 (2000 µg/ml) into columns 2 to 8.

    10.

    Method will pause until user resumes it again. Set a timer for 5 minutes.

    11.

    After 5 min, click OK to resume method.

    12.

    Transfer 200 µl from Buffer B to the 3 Read plates and incubate for 0h 10m 0s .

    The method will pause until user resumes it again. Set up a 10 minutes timer and click OK afterwards to finish.

    Spectrophotometer reading for protein Quantification

    13.

    Transfer plates to the microplate reader (MD Spectramax 250) to read absorbance at 750 nm and calculate protein concentration.

    14.

    Read Control Read plate.

    ABC
    SampleConcentrationMean OD Value (Absorbance)
    St011250.024
    St022500.042
    St035000.090
    St047500.138
    St0510000.165
    St0615000.239
    St0720000.281

    Standards (µg/ml)

    Example Standards Curve
    Example Standards Curve
    15.

    Read Protein Read plate 1 and Protein Read plate 2.

    AB
    SampleConcentration
    Un88585.249
    Un89785.257
    Un90670.135
    Un91718.864
    Un92868.962
    Un93679.907
    Un94743.064
    Un95994.173
    Un961115.072

    Examples of Sample concentrations that are reported by MD Spectramax 250

    Note
    Remember to multiply the protein concentrations by five (5) to account for the five-fold dilution in Step #7.

    16.

    Store protein plate at -20°C until ready for Automated Protein Normalization and Tryptic Digestion on a Biomek-NX Liquid Handler System.