    Untargeted Top-down Proteomics by LC-MS/MS on Lumos

    Jeannie Camarillo, Bryon Drown, Neil Kelleher

    Published: 2022-03-01 DOI: 10.17504/protocols.io.bttnnnme
    AI 解读
    浏览 1327

    Abstract

    Describes the LC-MS/MS data acquisition procedure for top-down proteomics samples using the Thermo Scientific Orbitrap Fusion Lumos Tribrid mass spectrometer

    Steps

    1.

    Samples were analyzed on a Thermo Scientific Orbitrap Fusion Lumos Tribrid mass spectrometer in line with a Dionex Ultimate 3000 RSLCnano system

    2.

    Samples (6µL ) were injected via the autosampler and loaded onto a self-packed trap column (150 μm i.d. x 2 cm length packed with PLRP-S 5-μm particles 1,000-Å pore size) for 0h 10m 0s with 100% loading buffer (94.8% water:5% acetonitrile:0.2% formic acid)

    3.

    Following a valve switch and initiation of the nanopump at 300 nL/min (buffer A: 94.8 % water, 5 % acetonitrile, 0.2 % formic acid; buffer B: 4.8 % water, 95 % acetonitrile, 0.2 % formic acid), proteins were separated on a self-packed analytical column (75 μm i.d. x 25 cm length packed with PLRP-S 5-μm particles 1,000-Å pore size) according to the following gradient for fractions 1-4:

    ABC
    Time (min)%BValve Position
      0  510_1
    10  51_2
    1315
    7045
    7295
    7695
    80  5
    90  5

    For fraction 5 and later, nanopump used the following gradient:

    ABCDEFGHIJKLMNOPQRS
    Time (min)%BValve Position
    0510_1
    1051_2
    1315
    7050
    7295
    7695
    805
    905
    4.

    Eluted proteins were ionized in positive ion mode nanoelectrospray ionization (nESI) using a pulled tip nanospray emitter (15-μm i.d. ×125 mm) packed with 1mm of PLRP-S 5-μm particles 1,000-Å pore size with a custom nano-source (http://proteomicsresource.washington.edu/docs/protocols05/UWPR_NSI_Source.pdf).

    AB
    High-High
    Spray voltage1800
    Sweep gas0
    Ion transfer tube temp320
    Application modeIntact Protein
    Pressure modeLow Pressure
    Advanced Peak DeterminationTrue
    Default charge state15
    S-lens RF30
    Source fragmentation15 eV

    Global MS parameters

    5.

    Precursor (intact protein) spectra were acquired at 120k FTRP

    AB
    High-High
    Detector typeOrbitrap
    Resolving power120000
    m/z RP measured200 m/z
    Scan range600-2000
    Mass rangeNormal
    AGC target1000000
    Normalized AGC target250%
    Max Injection Time100 ms
    Microscans4
    Data typeProfile
    PolarityPositive
    Use wide quad isolationTrue

    Parameters for MS1 acquisition

    6.

    The mass spectrometer was operated using a Top2 data-dependent acquisition mode. Precursor ions were filtered by intensity, charge state, and dynamic exclusion.

    AB
    Intensity minimum20000
    Intensity maximum1E20
    Included charge states6-60
    Include undetermined charge statesFalse
    Dynamic exclusion after n times1
    Dynamic exclusion duration60 s
    Mass tolerance1.5 m/z
    Exclude isotopesTrue

    Precursor selection filters for DDA

    7.

    Ions for fragmentations were isolated and fragmented via higher energy dissociation (HCD)

    AB
    High-High
    Detector typeOrbitrap
    Isolation modeQuadrupole
    Resolving power60000
    m/z RP measured200 m/z
    Scan range350-2000
    AGC target1000000
    Normalized AGC target2000%
    Max injection time400 ms
    Microscans4
    Isolation window3 m/z
    Activation typeHCD
    Collision energy27
    Collision energy modeFixed
    PolarityPositive

    Parameters for MS2 acquisition